| Downstream occlusion | Closed tap/clamp, kink, positional or blocked cannula, viscous fluid, low threshold | Patient → line → clear cause; release pressure before opening | Occlusion pressure & time on analyzer; sensor calibration |
| Upstream occlusion | Roller clamp closed, empty bag, vented/non-vented spike issue, kinked line above pump | Open clamp, check bag and spike | Upstream sensor test |
| Air in line | Bubbles, empty bag, out-gassing, wet/dirty sensor, set not seated | Stop, clamp, remove air, re-seat | Known-volume air test; clean sensor |
| Plunger / syringe not engaged | Grippers not closed, flange out of slot, damaged syringe | Re-load syringe correctly | Switch and gripper function |
| Near end / VTBI done → KVO | Planned events | Prepare next bag/syringe | — |
| Low / depleted battery | Unplugged, aged pack, charger fault | Connect mains | Run-time test; replace pack |
| Door open | Door not latched, worn latch | Close and re-check set | Latch and switch |
| System error / malfunction | Processor disagreement, motor fault, memory error | Replace pump; tag out; keep the log | Event log, self-tests, board/mechanism service |
| Inaccurate delivery | Wrong set or syringe selection, worn mechanism, height change, free flow | Report; check set and settings | Analyzer flow test |